Designed for Stable and Repeatable Environmental Conditioning
The chamber is designed to maintain a uniform temperature and humidity environment for long-term conditioning, stability testing, and controlled environmental exposure.
Precision Temperature & Humidity Control
The programmable PID system continuously monitors and adjusts chamber conditions for stable operation.
Temperature fluctuation: ≤ ±0.5°C
Temperature uniformity: ≤ ±2.0°C
Humidity fluctuation: ≤ ±2.5% RH
Humidity range: 20%–98% RH
Uniform Air Circulation
Optimized airflow distributes conditioned air throughout the chamber, helping maintain consistent temperature and humidity around different samples.
Stable Humidification
The humidification system provides precise moisture control for maintaining the required relative humidity during extended conditioning.
Reliable 24/7 Operation
Heating, refrigeration, humidification, and air circulation systems are designed for continuous long-term operation.
Programmable Touchscreen Controller
The touchscreen controller supports temperature and humidity setting, multi-step programming, real-time monitoring, data logging, and alarm management.
Constant Temperature Humidity Conditioning Chamber Specifications
Model | TH-100 | TH-225 | TH-500 | TH-1000 |
Internal Dimension (mm) | 400*500*500 | 500*600*750 | 700*800*900 | 1000*1000*1000 |
Overall Dimension (mm) | 900*1050*1620 | 1000*1140*1870 | 1200*1340*2020 | 1500*1540*2140 |
Interior Volume | 100L | 225L | 500L | 1000L |
Heat load | 1000W |
Temperature Range | A : -20℃ ~ +150 ℃ B : -40℃ ~ +150 ℃ C: -70℃ ~ +150 ℃ |
Temperature Fluctuation | ± 0.5 ℃ |
Temperature Deviation | ± 2.0 ℃ |
Humidity Range | 20% ~ 98% RH |
Humidity Deviation | ± 2.5% RH |
Cooling Rate | 1 ℃ / min |
Heating Rate | 3 ℃ / min |
Controller | Programmable color LCD touch screen controller,PC Link, LAN interface for remote control |
Cooling system | Mechanical compression refrigeration system,environmentally friendly refrigerant,compliant with the GWP standards requirements |
Water supply system | Automatic water supply; Water purification system |
Safety device | Humidifier dry-combustion protection; over-temperature protection; over-current protection; Refrigerant high-pressure protection; Water shortage protection, Earth leakage protection; |
Interior Material | SUS304 stainless steel |
Exterior Material | Steel Plate with protective coating |
Key Features & Engineering Details
Smart Color Touchscreen ControllerMulti-language interface with high-precision graph display. One-touch startup for standard conditioning points (25°C/60%RH & 40°C/75%RH). Automatic USB data logging for audit compliance.
|
|
Precision Steam Humidification System |
|
Ergonomic & Laboratory-Friendly DesignMulti-tier adjustable stainless steel SUS304 shelves for flexible sample positioning. Heated glass viewing window with interior LED lighting for non-disruptive sample monitoring. Low-vibration, silent compressor system suitable for clean laboratory environments.
|  |
FAQs on Constant Temperature Humidity Conditioning Chamber
What are common conditioning conditions?
Common conditions include 25 ± 2°C / 60 ± 5% RH, 30 ± 2°C / 65 ± 5% RH, and 40 ± 2°C / 75 ± 5% RH. The required condition depends on the applicable standard, product specification, or testing protocol.
Can the chamber be used for ICH stability testing?
Yes. The chamber can be configured for temperature and humidity conditions commonly used in ICH Q1A(R2) stability studies, including 25°C/60% RH, 30°C/65% RH, and 40°C/75% RH.
Can it operate continuously for long-term conditioning?
Yes. The chamber is designed for continuous operation and can maintain programmed temperature and humidity conditions over extended periods for conditioning and stability studies.
Can the temperature and humidity conditions be customized?
Yes. LIB industry can customize the temperature range, humidity range, chamber volume, shelving, access ports, controller functions, and other configurations according to specific testing requirements.